Product Name: Total Residual Chlorine Online Analyzer
Product Model: RCl900
System Overview
Total residual chlorine refers to the amount of residual chlorine remaining after chlorine-based oxidants are added to water and have been in contact for a period of time, excluding the amount consumed by microorganisms, organic and inorganic substances in the water. It includes free residual chlorine and combined residual chlorine. Free residual chlorine refers to chlorine existing in the form of hypochlorous acid (HClO) or dissolved elemental chlorine (Cl2). Combined residual chlorine refers to chlorine existing in the form of chloramines and organochloramines. The presence of residual chlorine in water environments is mainly introduced by artificial chlorination treatment. Chlorination treatment is a water treatment technology that uses the strong oxidizing effect of chlorine to non-selectively kill bacteria, algae and other microorganisms in water. With the extensive use of chlorine, the toxic and carcinogenic effects of its by-products have increasingly drawn attention, making it necessary to analyze and determine the chlorine content. At the same time, to achieve precise dosing and economically and effectively exert the disinfection and sterilization effect of chlorine, it is also necessary to monitor the residual chlorine content in water in real time.
The RCl900 total residual chlorine online analyzer is a fully automatic online monitor controlled by an embedded system. After the water sample and the anti-interference agent are automatically injected into the reactor, the characteristic color-developing agent is automatically added for the color-developing reaction, and the absorbance of the reactants is measured. The online analyzer can directly measure the total residual chlorine content in water through the absorbance value.
System Features
- The water sample pretreatment device adopts a maintenance-free design, which can ensure that the maintenance cycle of the pretreatment device exceeds half a year.
- Undisturbed, suitable for various water bodies, capable of real-time monitoring of total residual chlorine, in compliance with standards from various industries such as environmental protection, medical care, and power, and can meet environmental protection acceptance requirements.
- Fully imported components and innovative analytical flow path design and reagent formula ensure extremely high measurement reproducibility.
- It operates fully automatically without the need for personnel to stand by, and can achieve intelligent functions such as automatic zero adjustment, automatic calibration, automatic measurement, automatic cleaning, automatic maintenance, self-protection, and automatic recovery.
- The online monitoring methods are diverse, enabling manual measurement at any time, automatic timed measurement, and automatic periodic measurement.
- Simple operation and maintenance, no professional training required.
- Flexible detection, short measurement cycle, fast response speed, the detection process can be freely set, and the measurement cycle can be customized by the user.
- Automatically conduct quality control and calibration to ensure the consistency and reliability of test results.
Technical Parameters
| Measurement Principle | Colorimetric method |
| Measurement Range | 0-10mg/L |
| Measurement Methods | Timed, equal-interval, manua |
| Measurement Accuracy | 10% |
| Detection Limit | 0.01mg/l |
| Reproducibility | 5% |
| Signal Output | 4-20mA, RS232, RS485 |
| Calibration Method | Automatic timed calibration or manual calibration |
| Internal Sampling of the Instrument | An injection pump is used |
| External Sampling of the Instrument | Two methods are provided, namely submersible pump and self-priming pump |
| Display | 8-inch color touch screen, resolution 800*600 |
| DataStorage | One year of valid data |
| Water Sample Pressure | Below 0.5KG |
| Ambient Temperature | 0-40℃ |
| Power Consumption | Approximately 100W |
| Power Supply | 220VAC±10%/50-60Hz |
| Dimensions | 500x400x140mm |
| Weight | Approximately 20KG |
| Installation Method | Wall-mounted installation |