Product Name: Online Analyzer for Total Residual Chlorine
Product Model: RCl900
System Overview
Total residual chlorine refers to the amount of residual chlorine remaining in water after chlorine-based oxidants have been added and come into contact for a period of time, excluding the amount consumed by microorganisms, organic and inorganic substances in the water. It includes free residual chlorine and combined residual chlorine. Free residual chlorine refers to chlorine existing in the form of hypochlorous acid (HClO) or dissolved elemental chlorine (Cl2). Combined residual chlorine refers to chlorine existing in the form of chloramines and organochloramines. The presence of residual chlorine in water environments is mainly introduced by artificial chlorination treatment. Chlorination treatment is a water treatment technology that uses the strong oxidizing effect of chlorine to non-selectively kill bacteria, algae and other microorganisms in water. With the extensive use of chlorine, the toxic and carcinogenic effects of its by-products have increasingly attracted attention, and it is necessary to analyze and determine the chlorine content. Meanwhile, to achieve precise dosing and economically and effectively exert the disinfection and bactericidal effects of chlorine, it is also necessary to monitor the residual chlorine content in water in real time.
The RCl900 total residual chlorine online analyzer is a fully automatic online monitor controlled by an embedded system. After the water sample and the anti-interference agent are automatically injected into the reactor, the characteristic chromogenic agent is automatically added for the color reaction, and the absorbance of the reactants is measured. The online analyzer can directly measure the total residual chlorine content in water through the absorbance value.
System Features
· The water sample pretreatment device adopts a maintenance-free design, which can ensure that the maintenance cycle of the pretreatment device exceeds half a year.
· It is undisturbed, suitable for various water bodies, and can monitor total residual chlorine in real time. It complies with standards of various industries such as environmental protection, medical care, and power, and can meet environmental protection acceptance requirements.
· All imported components, innovative analytical flow path design and reagent formula ensure extremely high measurement reproducibility.
· It operates fully automatically without the need for personnel to stand by. It can achieve intelligent functions such as automatic zero adjustment, automatic calibration, automatic measurement, automatic cleaning, automatic maintenance, self-protection and automatic recovery.
· The online monitoring methods are diverse, enabling manual measurement at any time, automatic timed measurement, and automatic periodic measurement.
· Simple operation and maintenance, no professional training required;
· Flexible detection, short measurement cycle, fast response speed, the detection process can be freely set, and the measurement cycle can be customized by the user.
· Automatically carry out quality control and calibration to ensure the consistency and reliability of test results.
Technical Parameters
Measurement principle: Colorimetric method (meeting the requirements for comparison and acceptance);
Measurement range: 0-10mg/L;
Measurement methods: Timed, equal interval, manual;
Measurement accuracy: 10%;
Detection limit: 0.01mg/l;
Reproducibility: 5%;
Signal output: 4-20mA, RS232, RS485;
Correction method: Automatic timed correction or manual correction;
Internal sampling of the instrument: An injection pump is adopted;
External sampling of the instrument: Two methods are provided: submersible pump and self-priming pump;
Display: 8-inch color touch screen, resolution 800*600;
Data storage: One-year valid data;
Water sample pressure: less than 0.5KG;
Ambient temperature: 0-40℃;
Power consumption: Approximately 100W;
Power supply: 220VAC±10%/50-60Hz;
Size: 500x400x140mm;
Weight: Approximately 20 kilograms;
Installation method: Wall-mounted installation.